SIREMOBIL Iso-C
Maintenance Instructions
System
SIREMOBIL Iso-C
The protocol SPR2-230.105.02.06.02 is required for
these instructions
Print No.:
SPR2-230.101.02.06.02
Replaces: SPR2-230.101.02.05.02
02829450
SP
1998
© Siemens AG
The reproduction, transmission or use
of this document or its contents is not
permitted
without
express
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
English
Doc. Gen. Date: 07.05
written

Advertisement

Table of Contents
loading

Summary of Contents for Siemens SIREMOBIL Iso-C

  • Page 1 A r n o l d SIREMOBIL Iso-C Maintenance Instructions System SIREMOBIL Iso-C The protocol SPR2-230.105.02.06.02 is required for these instructions 02829450 1998 © Siemens AG The reproduction, transmission or use of this document or its contents is not permitted...
  • Page 2 Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
  • Page 3: Table Of Contents

    Electrical safety ............32 Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 3 of 46 Medical Solutions 07.05...
  • Page 4 Final Work Steps............45 6 _______ Changes to Previous Version _____________________________________ 46 SIREMOBIL Iso-C SPR2-230.101.02.06.02...
  • Page 5: Prerequisites

    If the system is equipped with 3D reconstruction, in combination with a navigation system, the required maintenance has to be per- formed together with a navigation system engineer. Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 5 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 6: Required Documents

    Systems equipped with an I.I. laser light localizer • Installation and Setting Instructions RXR2-130.033.03 1.2.5 Systems equipped with a laser light localizer • Setting Instructions SPR2-230.032.01.. SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 6 of 46 07.05 CS PS 24 Medical Solutions...
  • Page 7: Systems Equipped With Dicom Bridge

    • DICOM Bridge Installation and Setting Instructions SPR2-130.033.04 1.2.7 Systems equipped with Codonics printer • Codonics printer EP 1660 Installation and Setting Instruc- SPR2-130.814.01 tions Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 7 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 8: Required Tools, Measurement And Auxiliary Devices

    Calibration phantom (if 3D recon- 75 51 620 G5486 struction option is available) • Service PC • Cable for connecting service PC to 99 00 440 RE999 host SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 8 of 46 07.05 CS PS 24 Medical Solutions...
  • Page 9 (see logbook) • Sealing compound 34 43 009 • Special oil (Optimol Optipit) 55 07 525 • Special oil (Slic Pac PTFE) 55 07 517 Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 9 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 10: Spare Parts Which May Be Needed

    Cable deflector wheels small/SIREMOBIL stand • Conductive rubber/SIREMOBIL stand and monitor trolley • Touch-up paint 34 44 403 • Spray paint 84 27 734 RH999 SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 10 of 46 07.05 CS PS 24 Medical Solutions...
  • Page 11: Notes And Symbols

    Fig. 1: Safety Notes Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 11 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 12: Safety Information And Protective Measures

    TD00-000.860.01... must be complied with. CAUTION Electrical voltage! Non-compliance can result in material damage. When working on the system, ESD regulations must be observed. SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 12 of 46 07.05 CS PS 24 Medical Solutions...
  • Page 13: Radiation Safety Information

    To avoid burns, switch the system off prior to touching parts or components and allow at least 5 minutes of cool- ing. Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 13 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 14 Perform the required work with special care and atten- tion to detail. If needed, wear working gloves. SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 14 of 46 07.05...
  • Page 15: Safety Information - Risk Of Infection

    NOTE The eye is not in immediate danger. (blinking reflex of the eye). Nevertheless do not look directly into the laser beam. Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 15 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 16: Information On The Protective Conductor Resistance Test

    (according to DIN VDE 0751-1:2001-10, Fig. C3). Pos. 1 System Pos. 2 Application part (not available) Pos. 3 Measurement setup (integrated into measuring device) SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 16 of 46 07.05 CS PS 24 Medical Solutions...
  • Page 17: Information On Measuring The System Leakage Current

    The measuring procedure indicated in the report must be used. If the first measured value has to be newly determined (see note), a measuring procedure can be selected (direct measurement or differential measurement). Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 17 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 18 18)), the system must be insulated during the measurement and must not be touched. Fig. 4: Measuring circuit for direct measurement of the system leakage current in compliance with DIN VDE 0751-1:2001-10, Fig. C5 for protection class I. SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 18 of 46 07.05...
  • Page 19 2.5 mA is not exceeded. The evaluation is not necessary in the case of a new determination. File the report sheet in the system folder or log book. Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 19 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 20: Explanation Of Symbols

    KSK No. Customer-specific code Installed Volume Component Maintenance Unit The steps identified by these abbreviations are part of the maintenance certificate and should be checked off upon completion. SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 20 of 46 07.05 CS PS 24...
  • Page 21: Maintenance Interval

    SIE 3D reconstruction with navigation system (if present) Check the overall navigation accuracy • 6 months, however after 150 patients (approx. 300 3D scans) at the latest. Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 21 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 22: Technical Safety Checks (Tsc)

    SIM Monitors/displays 2.2 Are all cables and cable guides secure 3. Electrical safety, check: and without any visible damage? SIE Cables and plugs SIE Voltage discharge rubber SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 22 of 46 07.05 CS PS 24...
  • Page 23 3. Electrical safety, check: resistance SIE Protective conductor test 5.1 Final work steps, check: SIE Protective conductor test 3.2 Leakage current or equivalent leakage n.a. current measurement Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 23 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 24 5. Functional checks n.a. 5.1 Function of the Emergency Stop switch 3. Mechanical safety, check: Emergency Stop switch 3. Electrical safety, check: SIE 3D reconstruction SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 24 of 46 07.05 CS PS 24 Medical Solutions...
  • Page 25 (e.g., safety floor Otherwise: n.a. plate)? 5.7 Are the positions of the table and sys- n.a. tem reproducible (e.g., zero positions, layer height, etc.)? Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 25 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 26 Enter the results of the check. Have the operator, or a person authorized by the operator, sign the TSC report. Hand the TSC report over to the operator, or a person authorized by the operator. SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 26 of 46 07.05...
  • Page 27: Visual Inspection Of Exterior

    Inspect the entire system for damage, such as damage to the housing or paint. Damage to the mains connection • Check power lines, plugs, socket panels and outlets for damage. Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 27 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 28: Inspection Of Environment

    Mains voltage • Measure the line voltage and compare it to the line voltage label on the monitor trolley. Internal line impedance • Measure the internal line impedance. SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 28 of 46 07.05 CS PS 24...
  • Page 29: Mechanical Safety

    Press the Emergency Stop switch. The lifting column may not be raised or lowered. • Press the Emergency Stop switch again. The lifting column can be moved again. Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 29 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 30: I.i. Laser Light Localizer

    The guide rails must not be damaged. • The guide rails must lock properly into place when fully pushed into position. • If necessary, use a cleaning brush to remove dirt from the rails. SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 30 of 46 07.05...
  • Page 31: Navigation System

    - Replace any illegible labels or markings. ID labels • Ensure that all required ID labels are attached and in good condition. - Replace any illegible labels or markings. Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 31 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 32: Electrical Safety

    10 minutes and then every 5 or 10 minutes after that. Dose rate • Switch the system on • Check the dose rate (refer to the service instructions for SIREMOBIL ISO-C). Iris collimator SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 32 of 46 07.05...
  • Page 33: Monitor(S)/Displays

    Area-dose product measuring device • Check the area-dose product measuring device • (refer to the service instructions for SIREMOBIL ISO-C). Codonics printer cable break • Inspect all connection lines between the printer and the Framegrabber for damage and breaks. Inspect the cable routing of the energy chain for damage and proper functional- ity (ensure that none of the cables are pinched).
  • Page 34: Operation

    Check whether the operating icons and button labels are clearly legible and recogniz- able. • Check whether the light indicators of the operating elements and displays are properly functioning. • Check all system operating functions. SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 34 of 46 07.05 CS PS 24 Medical Solutions...
  • Page 35: Reconstruction

    44), IQ quick test. • Check the overall accuracy of the navigation system together with an engineer of the navigation system manufacturer. See "IQ quick test". Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 35 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 36: Conductive Rubber

    Monitor trolley voltage discharge rubber • Inspect the voltage discharge rubber on the monitor trolley for damage or dirt. If neces- sary, clean or replace it. SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 36 of 46 07.05...
  • Page 37: Measurements

    The leakage current test is part of the safety inspection. However, it is recommended to perform it only at the end of maintenance, with the system closed. See (Final Work Steps / p. 45). Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 37 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 38: Maintenance

    In the event that malfunctions occur or image quality worsens due to insuffi- cient cleaning, please refer the customer to the maintenance activ- ities/interval described in the Codonics user manual. SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 38 of 46 07.05...
  • Page 39 Clean the printer as described in the maintenance chapter of the Codonics user manu- al. Use the customer’s cleaning pads (some cleaning pads are included in the delivery volume of the printer). Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 39 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 40: Operating Value Inspection

    40 Maintenance, Operating Value / Functional Inspection Operating value inspection Error memory • Read out the error memory of the system and analyze the results. SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 40 of 46 07.05 CS PS 24 Medical Solutions...
  • Page 41: Functional Inspection

    The slot diaphragm plates are no longer visible on the screen. • Select I.I. zoom format. • Release fluoroscopy briefly. The slot diaphragm plates are no longer visible on the edges of the image. Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 41 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 42 The previously indicated monitor display of position and angle of rotation must coincide with the actual position of the slot diaphragm. • Open the slot diaphragm completely and rotate it again. SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 42 of 46 07.05...
  • Page 43 NOTE Perform this check with the lowest possible kV values so that the front edges of the slot diaphragm plate are effectively visualized. Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 43 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 44 Final Result / Quality Inspection and General Main- tenance Image quality (IQ) quick test • Perform the IQ quick test according to the SIREMOBIL ISO-C image quality quick test specifications. • Include additional monitors and displays in the test. DICOM bridge IQ test •...
  • Page 45: Final Work Steps

    (damaged insulation, damage caused by water ingress or humidity, defective interference suppressor, etc.) - even if the limit value of 2.5 mA is not exceeded. Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C Page 45 of 46 Medical Solutions 07.05 CS PS 24...
  • Page 46 Changes to Previous Version Changes to Previous Version Document was converted to DMS SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG Page 46 of 46 07.05 CS PS 24 Medical Solutions...