s
English
SITRANS F M MAGFLO
Electromagnetic flowmeter
2 kB SENSORPROM
memory unit
Order
no.: FDK:521H0725
INSTRUCTIONS
SITRANS F M MAGFLO
2 kB SENSORPROM
The enclosed SENSORPROM
type MAG 5000, MAG 6000, MAG 2500 and MAG 3000 manufactured week 47/1997
or later.
Transmitters type MAG 2500 and MAG 3000 manufactured before week 47/1997
require another type of SENSORPROM
Alternatively, the calibration factor can always be entered manually through the
operator menu of the transmitter. The flowmeter can then be used without a
®
SENSORPROM
memory unit installed.
This page has been updated (new drawing) 2004.01
Electromagnetic flowmeter
memory unit
®
memory unit (2 kB) must be used with transmitters
®
memory unit
Standard compact installation
The SENSORPROM
as standard delivered mounted in the
terminal box of the sensor. The connec-
tions to the transmitter are made when
the connection plate is located in the
terminal box.
Remote installation
When a wall mounting unit is used for
remote installation, the SENSOR-
®
PROM
unit must be removed from the
sensor and installed at the bottom of the
connection plate or, if possible, directly
in the wall mounting unit.
MAG 2500 and MAG 3000 Compact,
Mk I, II & III
The SENSORPROM
installed as shown below. Connector
and fixing screws are supplied with
transmitters manufactured week 7/1998
or later.
MAG 2500 and MAG 3000 19" insert
version
The SENSORPROM
stalled as shown below. Connector and
fixing screws are supplied with trans-
mitters manufactured week 7/1998 or
later.
®
memory unit is
®
unit must be
®
unit must be in-
SFIDK.PI.026.H2.52