OCTELL OT-100 Operation Manual

Forehead thermometer

Advertisement

Quick Links

OCTELL
OT 100
: OT 100

Advertisement

Table of Contents
loading

Summary of Contents for OCTELL OT-100

  • Page 1 OCTELL OT 100 : OT 100...
  • Page 2 Director Product introduction  Product structure description  Product display introduction  How to install the battery  Setting operation  Measurement method  Maintenance and precautions  Troubleshooting  Product specifications  Contraindications, precautions, warnings, and  instructions Graphics, symbols, abbreviations, etc. used in ...
  • Page 3: Product Introduction

    1.Product introduction This infrared forehead thermometer is a high-quality product. It uses infrared technology. Each time it is turned on, it will enter a self-test to ensure the accuracy of the measurement. This infrared forehead thermometer is mainly used to measure the temperature of the human forehead.
  • Page 4: Safety Instructions

    2.Safety instructions This product is only suitable for the purpose described in the instructions, such as damage caused by abnormal use. The manufacturer is not legally responsible. 2.1 Never put this product into water or other liquids. For cleaning, wipe with a damp soft cloth.
  • Page 5: Product Display

    4.Product display Test mode 2. Display value 3. Battery symbol 4. Storage location 5. Read out stored data 6. Temperature unit 7.Beep symbol 5.How to install the battery This machine comes with DC.3V (two AAA batteries). When the low voltage icon flashes on the screen, you need to replace the battery.
  • Page 6: Maintenance And Precautions

    Temperature Decrease Increase 38℃ alarm point 0.1 ℃ 0.1 ℃ range Temperature Offset Offset 0.1 ℃ -3 ℃ offset 0.1 ℃ 3 ℃ Beep switch turn off open open 7. Test method The infrared forehead thermometer has no display on the LCD screen in standby mode.
  • Page 7 10. Product Specifications Display range: 35.1°C~42.8°C(95.18℉~109.04℉) Temperature 0.1℃/32.18 °F resolution Measuring 1cm - 15cm 1cm-1.5cm 3CM~5CM distance Memory array 32 groups Memory Count Clinical ±0.3℃(±32.54℉) repeatability Maximum ±0.3℃(±32.54℉) allowable error Operating Temperature: 15 ℃ ~ 40 ℃ (9℉~104℉) environment Relative humidity: ≦ 85% Automatic 20 ±2 Seconds Shut-Down...
  • Page 8 12. Medical device graphics Reference instructions Note, refer to the attached file Keep dry B-type application equipment Avoid the sun Low battery alert Place up Fragile, handle with care 13. Electromagnetic compatibility related note: This product complies with the relevant electromagnetic compatibility requirements of YY0505 standard;...
  • Page 9 annex: Guidance and manufacturer's declaration-electromagnetic emissions This product is intended for use in the electromagnetic environment specified below. The purchaser or user of this product should ensure that it is used in this electromagnetic environment.Use: Launch test Compliance Electromagnetic Environment-Guide Radio frequency emission This product uses radio team 1...
  • Page 10 surge ± 1 kV Not applicable Not applicable GB / T 17626.5 differential mode voltage ± 2 kV common mode voltage <5% UT for 0.5 Temporary voltage cycles on power input line (On UT,> 95% dips) Drops, short 40% UT for 5 cycles Not applicable Not applicable interruptions and...
  • Page 11 3 V/m 3 V/m frequency of the Radio frequency 80 MHz~2.5 GHz transmitter. radiation Recommended isolation GB / T 17626.3 distance d = 1.2 P d = 1.2 P 80 MHz to 800 d = 2.3 P 800 MHz ~ 2.5 GHzWhere: P —...
  • Page 12 measured field strength of the product is higher than the applicable RF compliance level described above, the product should be observed to verify normal operation. If abnormal performance is observed, supplementary measures may be necessary, such as reorienting or repositioning the product. b In the entire frequency range from 150KHz to 80MHz, the field strength should be lower than 3 V / m.