Siemens SITRANS F C MASSFLO MASS 2100 Operating Manual

Siemens SITRANS F C MASSFLO MASS 2100 Operating Manual

Mass series flowmeters signal converter and sensor
Hide thumbs Also See for SITRANS F C MASSFLO MASS 2100:

Advertisement

s
SITRANS F C MASSFLO
MASS flowmeters
Signal converter type MASS 6000 19"
sensor type MASS 2100
Order no.: FDK:521H1179
SFIDK.PS.028.R1.02 -
A5E00253632
[
Operating Manual
    
*083R9507*
]

Advertisement

Table of Contents
loading

Summary of Contents for Siemens SITRANS F C MASSFLO MASS 2100

  • Page 1 Operating Manual      SITRANS F C MASSFLO MASS flowmeters Signal converter type MASS 6000 19" sensor type MASS 2100 Order no.: FDK:521H1179 *083R9507* SFIDK.PS.028.R1.02 - A5E00253632...
  • Page 2: Table Of Contents

         SITRANS F C MASSFLO Contents Introduction ........................3 Installation ........................4 Installation of signal converter MASS 6000 ..............4 Installation of sensor MASS 2100 ................... 4 Ex survey according to Directive 94/9/EC (ATEX) ............6 Overview and intrinsically safe data ................
  • Page 3: Introduction

    • Siemens Flow Instruments want to assist by estimating the chemical resistance of the sensor parts that are in connection with the media, but it is at any time the customer’s responsibility,...
  • Page 4: Installation

         SITRANS F C MASSFLO 2. Installation Installation Signal converter MASS 6000 can be installed in the safe area only! 2.1 Installation of signal converter MASS 6000 Safe area only! 2.2 Installation of sensor Category 1 equipment MASS 2100 Sensor MASS 2100 may be installed in zone 0, zone 1 and zone 2.
  • Page 5      2. Installation SITRANS F C MASSFLO Installation of sensor Mount the adaptor on top of the sensor inter- MASS 2100 (continued) face, if not already mounted. When fitting the multiple plug, please make sure that it is correctly oriented (note the little tap).
  • Page 6: Ex Survey According To Directive 94/9/Ec (Atex)

         SITRANS F C MASSFLO 2. Installation II 2G E Ex ia IIC T6 as an example 2.3 Ex survey according to Directive 94/9/EC (ATEX) Instrument groups Applies to instruments used in underground mining operations, as well as their above ground operations, which can be endangered by mine gas and/or flammable dusts.
  • Page 7: Overview And Intrinsically Safe Data

    [pF] Cable between MASS 2100 Maximum distance between sensor MASS 2100 and signal converter MASS 6000 is 300 m with and MASS 6000 Siemens Flow Instruments cable FDK:083H3005 or FDK:083H3006 or equivalent cable. Cable data Capacitance 300 [pF/m] 1 [µH/m]...
  • Page 8: Device Identification

         SITRANS F C MASSFLO 2. Installation 2.5 Device identification Sensor MASS 2100 label Signal converter MASS 6000 19" IP 20 and IP 65 label Meaning Notified body for QA supervision: UL International DEMKO A/S, Denmark Explosion protected ATEX Equipment Group and Protection Category Maximum medium temperature...
  • Page 9: Remote Installation Signal Converter Mass 6000

         SITRANS F C MASSFLO 2. Installation 2.6 Remote installation signal converter MASS 6000 Safe area only! 1. Fit the SENSORPROM ® unit on the connection board supplied with the signal converter. The SENSORPROM ® unit is supplied with the sensor. 2.
  • Page 10: Approvals

    19" rack with an ingress protection of at least IP 20. • For IP 65 Siemens Flow Instruments code number FDK:083F5037 can be used. • On the connection board are terminals marked „No wiring connections“. These terminals are not ...
  • Page 11: Electrical Connection

         SITRANS F C MASSFLO 3. Electrical connection Electrical connection 1. Connection terminals to: IP 20 • Power supply • 3.1 Electrical connection In- and output IP 20 & IP 65 2. Connection terminals to: MASS 2100 sensor Intrinsically safe terminals! It is an absolute requirement that the wires/ terminals of the intrinsically safe circuits can-...
  • Page 12: Signal Converter Mass 6000 & Sensor Mass 2100

         SITRANS F C MASSFLO 3. Electrical connection 3.2 Signal converter MASS 6000 & sensor MASS 2100 Installation 1) Mains supply 115 to 230 V a.c. from building installation Class II. A switch or circuit-breaker (max. 15 A) shall be included in the building installation. It must be in close proximity to the equipment and within easy reach of the OPERATOR, and it shall be marked as the disconnecting device for the equipment.
  • Page 13: Technical Data

         SITRANS F C MASSFLO 4. Technical data Technical data 4.1 Sensor MASS 2100. Versions DI 1.5, DI 3, DI 6, DI 15, DI 25, DI 40 Versions DI 1.5 DI 3 DI 6 DI 15 DI 25 DI 40 inch...
  • Page 14: Signal Converter Mass 6000

         4. Technical data SITRANS F C MASSFLO 4.2 Signal converter MASS 6000 MASS 6000 19" IP 20/IP 65 version Measurement of Mass flow [kg/s], volume flow [l/s], fraction [%], Brix, density [kg/m ], temperature [°C] Current output Current 0-20 mA or 4-20 mA...
  • Page 15: Output Characteristic Mass 6000

         SITRANS F C MASSFLO 4. Technical data 4.3 Output characteristic Output Bidirectional mode Unidirectional mode MASS 6000 characteristics 0-20 mA 4-20 mA Frequency Pulse output Relay Power supply off Power supply on Error relay No error Error Limit switch or 1 set point...
  • Page 16: Commissioning

         SITRANS F C MASSFLO 5. Commissioning Commissioning 5.1 Keypad and display layout Keypad The keypad is used to set the flowmeter. The function of the keys are as follows: TOP UP KEY This key (hold 2 sec.) is used to switch between operator menu and setup menu.
  • Page 17: Menu Build-Up

         SITRANS F C MASSFLO 5. Commissioning Menu build-up The menu structure of a specific type of signal converter is shown in a menu overview map. Details of how a specific parameter is set is shown in a menu detail map for the specific parameter. The menu structure is valid for the title and subtitle line only.
  • Page 18: Menu Overview

         SITRANS F C MASSFLO 5. Commissioning 5.3 Menu overview SFIDK.PS.028.R1.02...
  • Page 19: Menu Details

         SITRANS F C MASSFLO 5. Commissioning 5.4 Menu details Basic settings menu The decimal point for flowrate, totalizer 1 and totalizer 2 can be individually positioned. • open the respective window. • ensure that the cursor is positioned below the decimal point. Use the SELECT KEY •...
  • Page 20: Outputs Setting Menu

         SITRANS F C MASSFLO 5. Commissioning 5.5 Outputs setting menu Current output The current output must be set off when not used. Digital output Pulse Digital output Frequency SFIDK.PS.028.R1.02...
  • Page 21      5. Commissioning SITRANS F C MASSFLO Digital output Batch Relay output Error level (Also possible through digital output) Acceptance level is set in the basic settings menu. Error number (Also possible through digital output) Limit switch & direction switch (Also possible through digital output)
  • Page 22: Reset Mode

    The standard deviation (Zero Sigma) must be within a window, which is predefined by Siemens Flow Instruments. This window is called Sigma Limit. If the standard deviation is outside the window the following message is shown in the display: “Zero Sigma exceeds Sigma Limit”.
  • Page 23: Service

         SITRANS F C MASSFLO 6. Service Service Error Error text #Comment Outputs Input 6.1 List of error numbers Remedy text status status I1 - Power on Power on has activated Active Active I2 - Add-on module Applied A new module has been added to the system Active...
  • Page 24: Trouble Shooting Mass 6000

         SITRANS F C MASSFLO 6. Service 6.2 Trouble shooting Symptom Output Error Cause Remedy MASS 6000 signals code Empty display Minimum 1. Supply voltage 1. Check supply voltage 2. MASS 6000 defective 2. Replace MASS 6000 No flow signal Minimum 1.
  • Page 25: Ordering

    Standard ..........................................Density ..........................................Brix/Plato ........................................... Fraction flow (specified by customer), contact Siemens Flow Instruments ..................... Flow calibration Standard calibration included in sensor (3 flow x 2 points) ............................Standard calibration, matched pair (3 flow x 2 points; FDK:085F7351) ........................
  • Page 26: Build-Up Ordering - High Pressure Versions

    Standard ..........................................Density ..........................................Brix/Plato ........................................... Fraction flow (specified by customer), contact Siemens Flow Instruments ..................... Flow calibration Standard calibration included in sensor (3 flow x 2 points) ............................Standard calibration, matched pair (3 flow x 2 points; FDK:085F7351) ........................
  • Page 27: Signal Converter 19" Ex-Version

         SITRANS F C MASSFLO 7. Ordering 7.3 Signal converter 19" Description Version Supply voltage Code no. Symbol Ex-version MASS 6000 signal converter, 1 current output 115-230 V.a.c./ FDK:083H0202 [EEx ia] IIC IP 20 version for 19" rack 1 frq./pulse output 50/60Hz and panel mounting...
  • Page 28: Accessories For Mass 6000 & Mass 2100

         SITRANS F C MASSFLO 7. Ordering 7.4 Spare parts for MASS Description Code no. Symbol 6000 & MASS 2100 Wall mounting enclosure for MASS 6000 19" version IP 65 (21 TE) FDK:083F5037 Wall boxes Wall mounting enclosure for MASS 6000 19" version IP 65 (Without back plates/PCB) (42 TE) FDK:083F5038...
  • Page 29: Certificates

         SITRANS F C MASSFLO 8. Certificates 8.0 Download certificates Please go to http://www.siemens.com/flow 8.1 EU-declaration of conformity SFIDK.PS.028.R1.02...
  • Page 30: Ec Type Examination Certificate

         SITRANS F C MASSFLO 8. Certificates 8.2 EC type examination certificate SFIDK.PS.028.R1.02...
  • Page 31      SITRANS F C MASSFLO 8. Certificates 8.2 EC type examination certificate (continued) SFIDK.PS.028.R1.02...
  • Page 32 Suggestions for improvement registration of a utility model or design, are reserved. are always welcomed. Technical data subject to change without prior notice. Copyright © Siemens AG 2003 All Rights Reserved SIEMENS Order no.: FDK:521H1179 Flow Instruments A/S...