SITRANS F US SONOFLO
Step 1
    
*
*
Valid for standard 160 mm transducers.
With other transducer lengths, set in dimensions correspond-
ing to these lengths.
SFIDK.PS.029.Y1.02
Step 3
Before proceeding it is a good idea to
measure L
finished. By means of a plumb line the
ends of the transducer holders are
transpositioned to the upper guiding
beam.
L
max
uring tape on the upper guiding beam.
Enter the values in the measuring report.
The distance between the transducer
windows (L) is calculated based on L
as shown.
The standard transducer length is
160mm. If other lengths have been or-
dered, use the new values in the calcu-
*
lation.
of the sound track just
max.
. can now be measured with a meas-
max
7