Page 2
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho- rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
Remote control, TechSwitch (Option) ......4 - 5 Siemens-Elema AB SPR8-220.031.02 Page 3 of 4 Mobilett Plus/ E/ M/ HP Solna, Sweden Rev. 02 11.02...
Page 4
0 - 4 Contents This page intentionally left blank. Mobilett Plus/ E/ M/ HP SPR8-220.031.02 Page 4 of 4 Siemens-Elema AB Service Rev. 02 11.02 SPS-UD Solna, Sweden...
1 - 1 Purpose This document describes how to install a remote exposure switch system on a Mobilett Plus/ Plus E/ Plus M/ Plus HP. Important note After the installation is done, separate the last chapter, "the Supplement to the Instruc- tions for Use", from the document.
Checks and adjustments performed with radiation ON are identified by the radiation warning symbol During these types of adjustments, radiation protective clothing must be worn. Observe all MOBILETT safety guidelines. Mobilett Plus/ E/ M/ HP SPR8-220.031.02 Page 2 of 2 Siemens-Elema AB Service Rev.
Measure the discharge procedure between measuring points CAPPOS and CAPNEG on the D7 board (pos 6 / Fig. 1) with a digital multimeter. NOTICE For MOBILETT Plus HP only: Remove the cover in front of the batteries. Disconnect one of the blue battery connector pairs.
Page 8
Smoothen the sharp edges of the newly drilled holes with a file. SIDE VIEW TOP VIEW ± ± ± ± ± Fig. 2 Positioning of holes Mobilett Plus/ E/ M/ HP SPR8-220.031.02 Page 2 of 8 Siemens-Elema AB Service Rev. 02 11.02 SPS-UD Solna, Sweden...
Page 9
(pos 1 / Fig. 5). Module connector Remote sensor Fig. 5 Siemens-Elema AB SPR8-220.031.02 Page 3 of 8 Mobilett Plus/ E/ M/ HP Solna, Sweden Rev. 02 11.02 SPS-UD Service...
Page 10
Fasten the ground connector cable (pos 10 / Fig. 6). Make sure the washers are oriented as described in Fig. 6. Position the provided tie-mounts to the areas indicated in Fig. 6 at pos 5, 6 and 7. Mobilett Plus/ E/ M/ HP SPR8-220.031.02 Page 4 of 8...
When setting the exposure parameters use the lowest possible values for kV and mAs. NOTICE For MOBILETT Plus HP only: Reconnect the loose battery connector pair. ’ON’ LED (Green, LED4) Description: Illuminates when the X-ray unit is turned on.
Page 12
Make a series of exposures to be certain the remote system is functioning properly. Set the kV to 90 and 125 (133 if possible) and maximum mAs for the exposures. Mobilett Plus/ E/ M/ HP SPR8-220.031.02 Page 6 of 8...
(with the remote finder feature and low battery indicator) in the holder, and store the second spare (included in the delivery) in a safe place in the Radiology Department. Mobilett Plus HP Mobilett Plus/ Plus E/ Plus M Fig. 7 Placing of remote holder on Mobilett Siemens-Elema AB SPR8-220.031.02...
Page 14
2 - 8 Installation This page intentionally left blank. Mobilett Plus/ E/ M/ HP SPR8-220.031.02 Page 8 of 8 Siemens-Elema AB Service Rev. 02 11.02 SPS-UD Solna, Sweden...
Page 15
3 - 1 New components added for installation; ferrite core and ground connector cable. Image on page 2-4 updated. Safety notes revised. Siemens-Elema AB SPR8-220.031.02 Page 1 of 2 Mobilett Plus/ E/ M/ HP Solna, Sweden Rev. 02 11.02 SPS-UD Service...
Page 16
3 - 2 Changes to previous version This page intentionally left blank. Mobilett Plus/ E/ M/ HP SPR8-220.031.02 Page 2 of 2 Siemens-Elema AB Service Rev. 02 11.02 SPS-UD Solna, Sweden...
Vertical column Remote control (Option) Remote control (Option) Remote holder (Option) Remote holder (Option) Transport safeguard Transport safeguard Mobilett Plus/ Plus E/ Plus M Mobilett Plus HP Siemens-Elema AB SPR8-220.031.02 Page 1 of 6 Mobilett Plus/ E/ M/ HP Solna, Sweden Rev.
(option) in line-of sight. This includes any units behind a window or lead glass. If there are other mobiles in line-of-sight, power down those mobiles while operating the other. Mobilett Plus/ E/ M/ HP SPR8-220.031.02 Page 2 of 6...
The beeps will continue indefinitely until the remote control is located and put back in the remote holder. Siemens-Elema AB SPR8-220.031.02 Page 3 of 6 Mobilett Plus/ E/ M/ HP Solna, Sweden Rev. 02 11.02 SPS-UD Service...
Press again and hold to make the exposure Battery compartment (9-volt alkaline included) * These features are not implemented in the spare remote control (included in the kit) Mobilett Plus/ E/ M/ HP SPR8-220.031.02 Page 4 of 6 Siemens-Elema AB Service Rev.
180i; Remote Sensor built into the front arm of X-ray unit Remote control kit includes: Two remote controls (one spare), IR sensor with modular cable, PCB electronics package with flat cable Siemens-Elema AB SPR8-220.031.02 Page 5 of 6 Mobilett Plus/ E/ M/ HP Solna, Sweden Rev. 02 11.02 SPS-UD Service...
Page 22
4 - 6 Supplement to the Instructions for Use This page intentionally left blank. Mobilett Plus/ E/ M/ HP SPR8-220.031.02 Page 6 of 6 Siemens-Elema AB Service Rev. 02 11.02 SPS-UD Solna, Sweden...