List of compatible components ......2 - 10 3 _______Test certificate, MOBILETT Plus HP _______________________________ 3 - 1...
Page 4
0 - 4 Contents This page intentionally left blank. MOBILETT Plus HP SPR8-220.033.01 Page 4 of 4 Siemens-Elema AB Rev. 05 05.01 SPS-UD Solna, Sweden...
Training of customer support engineers Due to the technology, used in this equipment, the setup, service and maintenance is only allowed to be performed by a customer support engineer with a Work Permit for MOBILETT Plus HP. Text emphasis WARNING “Warnings”...
(works with the filter comparison method) • Protective ground/earth meter • Equivalent leakage current meter • Storage oscilloscope with ± 2.5 % accuracy • Dose meter MOBILETT Plus HP SPR8-220.033.01 Page 2 of 8 Siemens-Elema AB Rev. 05 05.01 SPS-UD Solna, Sweden...
- the product-specific safety information in the documents, - as well as the general safety information contained in ARTD Part 2. • Connect the MOBILETT Plus HP only to line voltage (receptacle), that corresponds to the installation requirements of VDE 0107 or country-specific regulations. •...
Remove the system upper cover. • Discharge the capacitors in M10 and M11 with S10 (D7) (refer to Fig. 1). • Wait 5 minutes; then remove the left, right and front covers. MOBILETT Plus HP SPR8-220.033.01 Page 4 of 8 Siemens-Elema AB Rev. 05 05.01...
Page 9
- Check whether the voltage increases. - Switch the system OFF, disconnect the line voltage cable and switch S10 on. Check whether the voltage decreases. Siemens-Elema AB SPR8-220.033.01 Page 5 of 8 MOBILETT Plus HP Solna, Sweden Rev. 05 05.01 SPS-UD...
Page 10
In this case, use the Service Instructions to guide you in trouble- shooting. Use caution when performing measurements at the capacitors in M10 and M11. Use only the specified measurement devices (350 V DC). MOBILETT Plus HP SPR8-220.033.01 Page 6 of 8 Siemens-Elema AB Rev. 05 05.01...
Required measurements can be partially transferred from the test protocol provided. Supply schedule 1 crate: L = 1360 mm B = 780 mm H = 1860 mm Weight: gross approx MOBILETT Plus HP 370 kg (Technical documents enclosed) Siemens-Elema AB SPR8-220.033.01 Page 7 of 8 MOBILETT Plus HP Solna, Sweden Rev.
Page 12
1 - 8 General This page intentionally left blank. MOBILETT Plus HP SPR8-220.033.01 Page 8 of 8 Siemens-Elema AB Rev. 05 05.01 SPS-UD Solna, Sweden...
Check the unit for cracks, fractures or other damages. • Make sure that neither the single tank nor the collimator has any mechanical defects which would impair the radiation protection. Siemens-Elema AB SPR8-220.033.01 Page 1 of 10 MOBILETT Plus HP Solna, Sweden Rev. 05 05.01 SPS-UD...
If the formation status is OK, the "Ready" light indicator lights up in approximately 2 minutes. Continue with chapter "Function check". If error message "CAL 19" or "Err 17" appear on the display, see chapter "Capacitor formation". MOBILETT Plus HP SPR8-220.033.01 Page 2 of 10 Siemens-Elema AB Rev. 05 05.01...
Use a screwdriver and adjust the speed by potentiometer on D101: - turn clockwise (CW) for slower speed. - turn counterclockwise (CCW) for faster speed. NOTICE This will effect the slow/fast speed simultaneously. Siemens-Elema AB SPR8-220.033.01 Page 3 of 10 MOBILETT Plus HP Solna, Sweden Rev. 05 05.01 SPS-UD...
If the formation is successfully finished the display will show: kVmAs CAP150 = 150 Volt. Duration: Approximately 100 minutes. • Switch "1" of S1 on D1 to "OFF" MOBILETT Plus HP SPR8-220.033.01 Page 4 of 10 Siemens-Elema AB Rev. 05 05.01...
• Check the function of the remote exposure switch system according to chapter functional check of "Installation Instructions: Installation of remote exposure switch system". Siemens-Elema AB SPR8-220.033.01 Page 5 of 10 MOBILETT Plus HP Solna, Sweden Rev. 05 05.01 SPS-UD...
Place the measurement detector on a table or other. Using the light field, set the detector in the beam projection, see fig. 4. • Set the SID to approx. 50 cm (20"). The SID may vary between different kV-meters. Fig. 4 MOBILETT Plus HP SPR8-220.033.01 Page 6 of 10 Siemens-Elema AB Rev. 05 05.01...
• Press lamp button to store value and leave program. • Switch "1" of S1 on D1 to "OFF". • Check the chosen values. Siemens-Elema AB SPR8-220.033.01 Page 7 of 10 MOBILETT Plus HP Solna, Sweden Rev. 05 05.01 SPS-UD...
Mark the four corners with coins and mark one of the corners with two coins, see fig. 6. • Set 52 kV, 2.0 mAs on the control panel. • Trigger an exposure. MOBILETT Plus HP SPR8-220.033.01 Page 8 of 10 Siemens-Elema AB Rev. 05 05.01...
The value "first measured" has to be entered in the test certificate at the end of the instal- lation and setting instructions. Siemens-Elema AB SPR8-220.033.01 Page 9 of 10 MOBILETT Plus HP Solna, Sweden Rev. 05 05.01 SPS-UD...
List of compatible components Component Part No. Single tank (new) 60 77 379 X037E Single tank (repaired) 61 85 172 X037E Collimator (new) 63 17 775 X037E MOBILETT Plus HP SPR8-220.033.01 Page 10 of 10 Siemens-Elema AB Rev. 05 05.01 SPS-UD Solna, Sweden...
Page 23
Test certificate, MOBILETT Plus HP 3 - 1 This test certificate is subdivided by analogy to installation and adjustment instruction. This test certificate shall be completed in duplicate, one copy remaining with the client, the other to be filed at the branch office.
Page 24
3 - 2 Test certificate, MOBILETT Plus HP Check for reproducibility Fluctuation coefficient C: ........Measurement device Type: ......... Serial No.: ......Date calibrated: ....Check the arm- and single tank movement Check the turn plate Check the brake Light field relative to the x-ray field (A+C)x100% = ......
Page 25
Fig.1 and Fig. 2 improved Chap. 2 Page 2 Information change in chapter "Charging of batteries". Page 3 Information change regarding motor drive speed. Siemens-Elema AB SPR8-220.033.01 Page 1 of 2 MOBILETT Plus HP Solna, Sweden Rev. 05 05.01 SPS-UD...
Page 26
4 - 2 Changes to previous version This page intentionally left blank. MOBILETT Plus HP SPR8-220.033.01 Page 2 of 2 Siemens-Elema AB Rev. 05 05.01 SPS-UD Solna, Sweden...