Siemens ARCADIS Avantic Installation And Start-Up Manual

Siemens ARCADIS Avantic Installation And Start-Up Manual

Option area dose measurement
Hide thumbs Also See for ARCADIS Avantic:

Advertisement

Quick Links

W ü r t h w e i n
ARCADIS Avantic
Installation and Startup
Option
Area Dose Measurement
Print No.:
SPR2-330.814.01.01.02
Replaces: n.a.
10048590
10047654
SP
2005
© Siemens AG
The reproduction, transmission or use
of this document or its contents is not
permitted
without
express
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
English
Doc. Gen. Date: 02.05
written

Advertisement

Table of Contents
loading

Summary of Contents for Siemens ARCADIS Avantic

  • Page 1 W ü r t h w e i n ARCADIS Avantic Installation and Startup Option Area Dose Measurement 10047654 10048590 2005 © Siemens AG The reproduction, transmission or use of this document or its contents is not permitted without express written authority.
  • Page 2 Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
  • Page 3: Table Of Contents

    Final work steps ............8 2 _______ Changes to Previous Version______________________________________ 9 Siemens AG SPR2-330.814.01.01.02 ARCADIS Avantic Page 3 of 10 Medical Solutions 02.05...
  • Page 4: General Information

    X-ray radiation are identified by the radiation symbol x . Radiation protective clothing must be worn while performing procedures marked with this symbol. Additional information These instructions only apply to the retrofitting of the dose meter to an ARCADIS Avantic system. Required parts •...
  • Page 5: With The Laser Light Localizer

    Use the switch on the measuring chamber to select the desired resolution (see the in- scription on the measuring chamber and the measuring chamber operating instruc- tions). • Reattach the POWERPHOS cover. Siemens AG SPR2-330.814.01.01.02 ARCADIS Avantic Page 5 of 10 Medical Solutions 02.05 CS SD 24...
  • Page 6: Programming And Calibrating The Area Dose Product Measuring Device

    Programming and calibrating the area dose product measuring device NOTE This calibration operation is used to adjust the area dose product measuring device to the ARCADIS Avantic system display. Programming • Start and log on to the service program. •...
  • Page 7: Checking The Programmed Value (Standard Resolution/Pediatrics)

    Click <Diamentor> and ">" under "Adjustment" for the next page in the Service menu. • Click "Click Here" in the service view and check the accuracy of the diamentor accord- ing to the instructions on the system. Siemens AG SPR2-330.814.01.01.02 ARCADIS Avantic Page 7 of 10 Medical Solutions 02.05 CS SD 24...
  • Page 8: Final Work Steps

    Use <Go> to start the burning operation for the following packages: SW-Settings.., As- piaSettings, Security-Settings, ExamSet, MainSystem. • Perform partial acceptance tests in accordance with country-specific regulations. • Check the calibration of the laser light localizer if applicable. ARCADIS Avantic SPR2-330.814.01.01.02 Siemens AG Page 8 of 10 02.05 CS SD 24...
  • Page 9 Changes to Previous Version 2Changes to Previous Version N/A; first version Siemens AG SPR2-330.814.01.01.02 ARCADIS Avantic Page 9 of 10 Medical Solutions 02.05 CS SD 24...
  • Page 10 Changes to Previous Version ARCADIS Avantic SPR2-330.814.01.01.02 Siemens AG Page 10 of 10 02.05 CS SD 24 Medical Solutions...