Page 2
The service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Sie- mens or one of its affiliates to provide such services.
PTW-DIADOS 97 17 612 Safety tester e.g. BENDER Unimet 1100 51 38 727 Service PC * Service software (refer to SIREMOBIL Iso-C logbook) * Connection cable, Service-PC - SIREMOBIL Iso-C e.g. 99 00 440 1 set of resolution tests e.g.
Prerequisites Information regarding this document These service instructions will assist you in adjusting the SIREMOBIL Iso-C for routine operation. Not every adjustment is required each time the system is serviced. Position designations The designations used in these instructions (left, right, front, and back) are explained in (Fig.
Page 28
Refer to (Safety information / p. 11). Disconnect the SIREMOBIL Iso-C line voltage plug. Open the back cover of the monitor trolley. Measure the on-site line voltage and line frequency. Adjust transformer T1 to the required line voltage and line frequency. Refer to the label on the switch-on assembly.
Checking the operating voltages Low voltages 3Checking the operating voltages Check the operating voltages for the SIREMOBIL Iso-C against the values in the table. from test point to test point Voltage Potentiometer D1.X31 (0V) D1.X30 (+5 V) + 5.0 V to 5.2 V M14.
Loading the operating software Control board D1, host 4Loading the operating software Connect the service PC to the serial service interface on SIREMOBIL Iso-C. Start the service program. Insert the "Parameter" diskette into drive A. Select the Download menu and start the download (ARC / Fil / Host).
(filament circuit learning program). Always per- form the kV offset adjustment prior to beginning the genera- tor adjustment. Connect the service PC to the serial interface of the SIREMOBIL Iso-C. Start the service program. Select the Diagnostic menu. Select the Monitoring menu.
The kV offset adjustment must be performed prior to starting the generator adjustment. NOTE If an error occurs during the generator adjustment, switch OFF the SIREMOBIL Iso-C and perform the kV offset and gen- erator adjustments again. NOTE As soon as the generator adjustment is started, press the radiation release button on the hand switch until the "Stop...
With the system on, connect the service PC to the SIREMOBIL Iso-C. Select "Test..." in the Diagnostic menu. Select cassette simulation in "Function Groups". Select cassette switch on in "Available Functions". The SIREMOBIL Iso-C is now in direct exposure mode. Select 81 kV with the kV+ / kV- keys.
The "Adjustment mode" parameter in the "Doserate adjustment" parameter group has been programmed to "semi-automatic". Preliminary steps Switch on SIREMOBIL Iso-C. Connect the service PC to the serial interface of the SIREMOBIL Iso-C. Start the service program. Select the "Diagnostics..."menu. Select the "Monitoring..."menu.
Radiation protection is required. Wear a lead apron. Attach the center cross in the center of the image intensifier input screen. Connect the service PC to the serial interface of the SIREMOBIL Iso-C. Start the service program. Select the "Adjustment, Calibration" menu.
Page 53
Close the status window by clicking the OK button. For the query "Do you want to reset the Unit", click "Yes". Switch the SIREMOBIL Iso-C off and on. Check the X-Iris again in survey format and zoom format. The collimator blades must still be visible on the edge of the image.
Then click "Set Value". Click "Put to Unit". After the SIREMOBIL Iso-C has booted, compare the 0° position of the slot diaphragm blades to the display on the monitor. If necessary, repeat the programming. Click "Save to file".
Page 63
The image contents will rotate clockwise. Connect the service PC to the serial interface of the SIREMOBIL Iso-C. Start the service program. Select the Adjustment, Calibrations menu. Answer the query "yes".
Programming Connect the service PC to the serial interface of the SIREMOBIL Iso-C. Start the service program. Select the Adjustment, Parameter menu. In the Combo box "Parameter Groups:", select "Diamentor" .
Prerequisites The T value is already programmed. See Programming. Connect the service PC to the serial interface of the SIREMOBIL Iso-C. Start the service program. Select the Adjustment, Calibrations menu. Click "Yes" to answer the query.
Adjustments / Programs Overview of SIREMOBIL Iso-C Parameters 5.14 Select the Moreinfo.hlp help file on the service PC for a short description of the parameters available on the SIREMOBIL Iso-C. NOTE If parameters have been changed and saved in the "Adjust- ments, Parameter"...
5.15 User Setup User Setup is selected with the CTRL + U key combination. The available functions are described in the SIREMOBIL Iso-C Operating Instructions. Technical Setup Technical Setup is selected with the CTRL + T key combination. Enter the password and confirm with [RETURN]...
Page 75
Yes: When pressing the image rotation key, the current image will be computerized and displayed. NO: The rotated image will not be computerized when pressing the image rotation key. For SIREMOBIL Iso-C always set to YES. Product name Allows you to change the MEMOSKOP name displayed dur- ing system booting.
Page 76
Function: Comments: Unit serial number Enter the serial number of the SIREMOBIL Iso-C when it is equipped with a DICOM bridge. The serial number is required to identify the images transferred. Please note that you have to omit the leading 0 when entering the serial num- ber.
When you replace the MEMOSKOP, it will be configured for 2 monitors. This configuration displays USER SETUP and TECHNICAL SETUP on the second monitor. Therefore, USER SETUP and TECHNICAL SETUP are not visible for SIREMOBIL Iso-C systems with only one monitor.
Exit TECHNICAL SETUP by pressing the "HOME" key on the MEMOSKOP keyboard. MEMOSKOP, configuration and country-specific language (User Setup) Start the MEMOSKOP service program and transfer the country-specific language to SIREMOBIL Iso-C via the service PC. NOTE After completing the MEMOSKOP configuration, save the parameters on a diskette.
Set the jumpers and switches according to the circuit diagram. Perform the download procedure for control board D1 (Host/Arc/Fill) NOTE For SIREMOBIL Iso-C 3D, perform the download with the software delivered with the unit or 3D upgrade kit (see log- book).
SIREPHOS housing. For SIREMOBIL Iso-C systems with I.I. 23 cm (9”), no addi- tional 0.1 mm Cu filter needs to be installed. It is sufficient to reduce the maximum air kerma rate by reducing the maxi-...
WARNING Electrical voltage! (Safety information / p. 11). Switch the SIREMOBIL Iso-C off and wait at least > 3 minutes until the high voltage in the system and in the I.I. mini-voltage supply dissipates. WARNING Risk of injury on mechanical parts! If these guidelines are not followed, minor to moderate injury, especially to the hands, can occur.
Page 106
User Setup. Otherwise set to "Yes". Unit serial number Program the serial number of the SIREMOBIL. Do not enter the leading zero. Product name Program "SIREMOBIL Iso-C". Default program change- Default: Yes. Depending on country-specific regulations able configure Yes or No.
Page 109
Otherwise set to "Yes". Unit serial number Program the serial number of the SIREMOBIL. Do not enter the leading zero. Product name Program "SIREMOBIL Iso-C". Default program change- Default: Yes. Depending on country-specific regulations able configure Yes or No. Create external disk Configure No.
Page 112
Configure 1. Program a different edge enhancement value upon customer request only. Software backward com- Configure No. patible Product name Program "SIREMOBIL Iso-C". Default program change- Default: Yes. Depending on country-specific regulations able configure Yes or No. Create external disk If the option "3D Reconstruction"...
Page 115
Setup. Otherwise set to "Yes". Unit serial number Program the serial number of the SIREMOBIL. Do not enter the leading zero. Product name Program "SIREMOBIL Iso-C". Software backward compati- Configure No. Default program changeable Default: Yes. Depending on country-specific regulations configure Yes or No.
Mouse, external mouse, and PC keyboard 6.20 NOTE Only applies to SIREMOBIL Iso-C 3D. The system comes with a mouse and an external mouse. Both are connected to the mouse interface of the 3D reconstruction PC by means of a mouse Y-switch. The key- board is connected to the keyboard interface of the 3D reconstruction PC.
Replacing boards and components Video switcher 6.21 Fig. 35: Fig. 36: NOTE Only applies to SIREMOBIL Iso-C 3D. NOTE Video switcher 1 ((3/Fig. 35 / p. 121)) is used to switch the video signal between the MEMOSKOP and the 3D recon- struction PC.
11). Switch the system power supply off and disconnect the power plug. NOTE Only applies to SIREMOBIL Iso-C 3D. In the monitor trolley, PC board D210 is installed above the line power filter ((Fig. 37 / p. 122)) . Disconnect the system from the power supply by pulling the line power plug plug out of the socket.
6.23 Fig. 38: NOTE Only applies to SIREMOBIL Iso-C 3D. Disconnect the system from the power supply by pulling the line power plug plug out of the socket. The syngo software and the backup/restore disk are kept in the monitor trolley logbook.
Replacing boards and components SIMOMED HM monitor 6.24 NOTE Only applies to SIREMOBIL Iso-C 3D. Switch the system off. Replace the monitor. Perform the monitor settings as described in the Startup Instructions, Chapter 7. Perform the IQ quick test. SIREMOBIL Iso-C SPR2-230.840.01.01.02...
3D reconstruction system electronics 6.25 Fig. 39: NOTE Only applies to SIREMOBIL Iso-C 3D. The 3D reconstruction system electronics (PC boards D200 including MCB3 and D310) is attached in front of PC board D1 in the basic unit ((Fig. 39 / p.
6.26 Fig. 40: Fig. 41: Fig. 42: NOTE Only applies to SIREMOBIL Iso-C 3D. Loosen the belt tensioner of the toothed-belt orbital drive ((1/Fig. 40 / p. 126)). Disconnect the motor connecting cables (motor cables + incremental encoder). Loosen the 2 fastening screws of the mounting plate (including the motor) ((1/Fig.
Fig. 43: Fig. 44: Fig. 45: Fig. 46: NOTE Only applies to SIREMOBIL Iso-C 3D. Loosen the belt tensioner of the toothed-belt orbital drive ((1/Fig. 45 / p. 128)). Disconnect the connecting cables of the orbital motor (motor cables + incremental encoder).
6.28 Fig. 47: Fig. 48: NOTE Only applies to SIREMOBIL Iso-C 3D. Move the C-arm orbitally into the -95° end position: II side actuated The orbital limit switch on the tube unit side is actuated. Take note of the order in which the cables of the actual value potentiometer are con- nected.
Fig. 49: Fig. 50: NOTE Only applies to SIREMOBIL Iso-C 3D. Move the C-arm orbitally into the -95° end position. The orbital limit switch on the tube unit side is actuated. Loosen the two fastening screws of the actual value potentiometer on the front of the coupling unit, but do not unscrew them ((1/Fig.
Replacing boards and components Angulation part 6.30 Fig. 51: NOTE Only applies to SIREMOBIL Iso-C 3D. NOTE When the angulation part is vertical, the markings on the two toothed belts must coincide (see (Fig. 51 / p. 134)). If the toothed wheel ((1/Fig.
Noncompliance can lead to medium to severe injuries and/or property damage. To replace the vertical lift brake, the SIREMOBIL Iso-C must be placed on its side. This requires 6 people. To prevent damage to the SIREMOBIL Iso-C or the floor, we recommend using several wool blankets or corru- gated cardboard or other suitable floor covering.
Risk of injury! Noncompliance can lead to medium to severe injuries and/or property damage. To replace the vertical lift motor, the SIREMOBIL Iso-C must be placed on its side. This requires 6 people. To prevent damage to the SIRE- MOBIL Iso-C or to the floor, we recommend using sev- eral wool blankets or corrugated cardboard, or other suitable floor covering.
Page 146
After the counterbalance weights have been reattached on both sides, move the C-arm orbitally from one end to the other several times. In the case of SIREMOBIL Iso-C 3D, also perform a scan. If the belt tension is still insufficient, repeat the procedure and move the belt to the next tooth to shorten it further.