Trigger Sensitivity Test - HP 1652B Getting Started Manual

Logic
Hide thumbs Also See for 1652B:
Table of Contents

Advertisement

Table
3-l.
Performance
Test Record (continued)
Test
Voltage
Measurement
Accuracy
Test
Limits
desults
Range
Chan 1
Chan2
10.0 v
31.758 - 38.242 V
5.0 v
15.878 - 19.122 v
2.0 v
6.935 - 7.065 V
1.0 v
3.174 - 3.826 V
500 mV
1.586 - 1.914 V
200 mV
633.2 - 766.8 mV
100 mV
315.6 - 384.4 mV
50 mV
156.8 - 193.2 mV
20 mV
61.52 - 78.48 mV
15 mV
43.14 - 56.86 mV
DC Offset
Accuracy
Test
Range
1.0 v
500 mV
200 mV
100 mV
19.50 - 20.50 v
9.758 - 10.242 V
4.878 - 5.122 V
1.948 - 2.052 v
Chan 1
Ghan 2
Bandwidth
Test
Down
<3dB at 100 MHz
Chan 1
Chan 2
Time
Measurement
Accuracy
Test
10 ns
9.4 to 10.6 ns
100 ns
99.4 to 100.6 ns
500 ns
499.4 to 500.6 ns
1 us
989.5 to 1.010 us
2 us
1.990 to 2.010 us
Chan 1
Chan 2
Trigger
Sensitivity
Test
Chan
1
Chan 2
2 V/div
0.4 div at 1 MHz
0.4 div at 100 MHz
200 mV/div
0.4 div at 1 MHz
0.4 div at 100 MHz
20 mVldiv
0.4 div at 1 MHz
0.4 div at 100 MHz
-
--
_--
Performance
Tests
3-56
HP 16528/1653B
Service
Manual