2 - 10
Repeat measurements
After a repair - if this becomes necessary due to the work performed - and basically
always after maintenance, the substitute leakage current has to be measured again.
The value measured must not exceed the permissible maximum value of 2 mA. In addi-
tion, the value first measured must not be exceeded by more than 50%.
The results have to be documented.
List of compatible components
Component
Single tank (new)
Single tank (repaired)
Collimator (new)
MOBILETT Plus HP
Installation and setting
Part No.
60 77 379 X037E
61 85 172 X037E
63 17 775 X037E
SPR8-220.033.01
Rev. 05
05.01
Page 10 of 10
Siemens-Elema AB
SPS-UD
2
2
Solna, Sweden