Note 11
9
If two multi-pod state analyzers are on, the qualifier pop-up menu will
show that only four pattern recognizers are available to each analyzer.
Pattern recognizers a-d and the range recognizer are assigned to the
first analyzer created, and pattern recognizers e-h go with the second
analyzer. In the Full Qualifier Specification pop-up there will be only
one OR gate and one set of pattern recognizers.
State Trace Menu
HP 16528/1653B
lo-4
Front-Panel Reference
With this Full Qualifier Specification pop-up, you specify a logical
combination of patterns or ranges as the qualifier. The pattern
recognizers are always partitioned into the groups of four shown. Only
one operator is allowed between the patterns in a group. Patterns in
uncomplimented form (a, b, etc.) can only be ORed.
The complements of patterns ( # a, # b, etc.) can only be ANDed.
For example, if the first OR field (gate) is changed to AND, all the
patterns for that gate are complemented, as shown below.
A
1
x
2
Full
Qualifier
Specification
Figure 1
O-4.
Complemented Patterns